ABSTRACT: Purpose
To propose and test a modified wideband late gadolinium enhancement (LGE) magnetic resonance (MR) imaging technique to overcome hyperintensity image artifacts caused by implanted cardiac devices.Materials and methods
Written informed consent was obtained from all participants, and the HIPAA-compliant study protocol was approved by the institutional review board. Studies in phantoms and in a healthy volunteer were performed to test the hypothesis that the hyperintensity artifacts that are typically observed on LGE images in patients with implanted cardiac devices are caused by insufficient inversion of the affected myocardial signal. The conventional LGE MR imaging pulse sequence was modified by replacing the nonselective inversion pulse with a wideband inversion pul