Precision and agreement of corneal power measurements obtained using a new corneal topographer OphthaTOP.


ABSTRACT:

Purpose

To evaluate repeatability and reproducibility of anterior corneal power measurements obtained with a new corneal topographer OphthaTOP (Hummel AG, Germany) and agreement with measurements by a rotating Scheimpflug camera (Pentacam HR, Oculus, Germany) and an automated keratometer (IOLMaster, Carl Zeiss Meditec, Germany).

Methods

The right eyes of 79 healthy subjects were prospectively measured three times with all three devices. Another examiner performed three additional scans with the OphthaTOP in the same session. Within one week, the first examiner repeated the measurements using the OphthaTOP. The flat simulated keratometry (Kf), steep K (Ks), mean K (Km), J0, and J45 were noted. Repeatability and reproducibility of measurements were assessed by within-subject s