Unknown

Dataset Information

Centrifugal microfluidic platform for ultrasensitive detection of botulinum toxin.


ABSTRACT: We present an innovative centrifugal microfluidic immunoassay platform (SpinDx) to address the urgent biodefense and public health need for ultrasensitive point-of-care/incident detection of botulinum toxin. The simple, sample-to-answer centrifugal microfluidic immunoassay approach is based on binding of toxins to antibody-laden capture particles followed by sedimentation of the particles through a density-media in a microfluidic disk and quantification by laser-induced fluorescence. A blind, head-to-head comparison study of SpinDx versus the gold-standard mouse bioassay demonstrates 100-fold improvement in sensitivity (limit of detection = 0.09 pg/mL), while achieving total sample-to-answer time of <30 min with 2-μL required volume of the unprocessed sample. We further demonstrate quantif

SUBMITTER: Koh CY 

PROVIDER: S-EPMC4303339 | biostudies-literature | 2015 Jan

REPOSITORIES: biostudies-literature