Interference of cerebrospinal fluid total protein measurement by povidone-iodine contamination.


ABSTRACT:

Background

A falsely high cerebrospinal fluid (CSF) total protein (TP) result measured by pyrogallol red (PGR) method was suspected to be caused by preparation of the collection site with povidone-iodine (PVP-iodine) solution.

Methods

CSF TP was evaluated for interference in samples with different final concentrations of PVP-iodine (up to 0.25% PVP and 0.025% iodine) or iodine alone (up to 0.025% iodine) using three methods: PGR, modified biuret and benzethonium chloride (BZTC). Interference exceeding ±20% of the baseline value is considered clinically significant according to the criterion defined by the College of American Pathologists.

Results

There were positive interference with the PGR method and negative inference for the BZTC method in CSF samples spiked with