ABSTRACT: Objective
To compare antimüllerian hormone (AMH) levels among three commercially available AMH immunoassays: AMH Gen II (Beckman Coulter), Ultrasensitive AMH (Ansh Labs), and picoAMH (Ansh Labs).Design
Cross-sectional.Setting
Academic reproductive endocrinology program.Patient(s)
90 newly diagnosed breast cancer patients before cancer treatment.Intervention(s)
None.Main outcome measure(s)
Proportion of detectable AMH levels by immunoassay, and comparability among assays.Result(s)
At a mean age of 38.1 years, the median (interquartile range) AMH level for the cohort was 0.92 [1.35] ng/mL for the Gen II assay, 1.68 [2.30] ng/mL for the Ultrasensitive assay, and 1.52 [2.41] ng/mL for the picoAMH assay. Significantly higher proportions of