Enhancing MALDI time-of-flight mass spectrometer performance through spectrum averaging.
Ontology highlight
ABSTRACT: Matrix-assisted laser desorption ionization time-of-flight (MALDI-TOF) mass spectrometers are simple and robust mass spectrometers used for analysis of biologically relevant molecules in diverse fields including pathogen identification, imaging mass spectrometry, and natural products chemistry. Despite high nominal resolution and accuracy, we have observed significant variability where 30-50% of individual replicate measurements have errors in excess of 5 parts-per-million, even when using 5-point internal calibration. Increasing the number of laser shots for each spectrum did not resolve this observed variability. What is responsible for our observed variation? Using a modern MALDI-TOF/TOF instrument, we evaluated contributions to variability. Our data suggest a major component of variabi