Validation of a microsphere immunoassay for serological leptospirosis diagnosis in human serum by comparison to the current gold standard.
Ontology highlight
ABSTRACT: A microsphere immunoassay (MIA) utilising Luminex xMap technology that is capable of determining leptospirosis IgG and IgM independently was developed. The MIA was validated using 200 human samples submitted for routine leptospirosis serology testing. The traditional microscopic agglutination (MAT) method (now 100 years old) suffers from a significant range of technical problems including a dependence on antisera which is difficult to source and produce, false positive reactions due to auto-agglutination and an inability to differentiate between IgG and IgM antibodies. A comparative validation method of the MIA against the MAT was performed and used to determine the ability of the MIA to detect leptospiral antibodies when compared with the MAT. The assay was able to determine samples in th
SUBMITTER: Wynwood SJ
PROVIDER: S-EPMC4373873 | biostudies-literature | 2015 Mar
REPOSITORIES: biostudies-literature
ACCESS DATA