End-tidal carbon dioxide monitoring using a naso-buccal sensor is not appropriate to monitor capnia during non-invasive ventilation.


ABSTRACT:

Background

In acute respiratory failure, arterial blood gas analysis (ABG) is used to diagnose hypercapnia. Once non-invasive ventilation (NIV) is initiated, ABG should at least be repeated within 1 h to assess PaCO2 response to treatment in order to help detect NIV failure. The main aim of this study was to assess whether measuring end-tidal CO2 (EtCO2) with a dedicated naso-buccal sensor during NIV could predict PaCO2 variation and/or PaCO2 absolute values. The additional aim was to assess whether active or passive prolonged expiratory maneuvers could improve the agreement between expiratory CO2 and PaCO2.

Methods

This is a prospective study in adult patients suffering from acute hypercapnic respiratory failure (PaCO2 ≥ 45 mmHg) treated with NIV. EtCO2 and expiratory CO2 v