ABSTRACT: Purpose
To develop, characterize, and implement a fast patient localization method for total marrow irradiation.Methods and materials
Topographic images were acquired using megavoltage computed tomography (MVCT) detector data by delivering static orthogonal beams while the couch traversed through the gantry. Geometric and detector response corrections were performed to generate a megavoltage topogram (MVtopo). We also generated kilovoltage topograms (kVtopo) from the projection data of 3-dimensional CT images to reproduce the same geometry as helical tomotherapy. The MVtopo imaging dose and the optimal image acquisition parameters were investigated. A multi-institutional phantom study was performed to verify the image registration uncertainty. Forty-five MVtopo images were
SUBMITTER: Takahashi Y
PROVIDER: S-EPMC4385424 | biostudies-literature | 2015 Jan