Model-based acceleration of look-locker T1 mapping.
Ontology highlight
ABSTRACT: Mapping the longitudinal relaxation time T1 has widespread applications in clinical MRI as it promises a quantitative comparison of tissue properties across subjects and scanners. Due to the long scan times of conventional methods, however, the use of quantitative MRI in clinical routine is still very limited. In this work, an acceleration of Inversion-Recovery Look-Locker (IR-LL) T1 mapping is presented. A model-based algorithm is used to iteratively enforce an exponential relaxation model to a highly undersampled radially acquired IR-LL dataset obtained after the application of a single global inversion pulse. Using the proposed technique, a T1 map of a single slice with 1.6mm in-plane resolution and 4mm slice thickness can be reconstructed from data acquired in only 6s. A time-consuming
SUBMITTER: Tran-Gia J
PROVIDER: S-EPMC4393277 | biostudies-literature | 2015
REPOSITORIES: biostudies-literature