Prolonged treatment time deteriorates positioning accuracy for stereotactic radiosurgery.


ABSTRACT:

Introduction

The accuracy of radiation delivery is increasingly important as radiotherapy technology continues to develop. The goal of this study was to evaluate intrafractional motion during intracranial radiosurgery and the relationship between motion change and treatment time.

Methods and materials

A total of 50 treatment records with 5988 images, all acquired during treatments with the CyberKnife Radiosurgery System, were retrospectively analyzed in this study. We measured translation and rotation motion including superior-inferior (SI), right-left (RL), anterior-posterior (AP), roll, tilt and yaw. All of the data was obtained during the first 45 minutes of treatment. The records were divided into 3 groups based on 15-min time intervals following the beginning of treatme