Accuracy and Reproducibility of High-definition Anorectal Manometry and Pressure Topography Analyses in Healthy Subjects.


ABSTRACT:

Background & aims

High-definition anorectal manometry (HDAM-3D) provides a topographic and 3-dimensional profile of anorectal pressure. We assessed anorectal sensorimotor function and the reproducibility of measurements made with HDAM-3D in healthy adults, and the accuracy of data analysis by its software.

Methods

Anal sphincter pressures and rectal sensory thresholds were measured in 78 healthy subjects via placement of a 10-mm rigid probe, with 256 circumferentially arrayed pressure sensors, and a balloon in the rectum. The bearing down maneuver was assessed in a subset of 18 subjects. We compared data analyzed by an expert with findings from automated software analysis. Measurements made in a subset of 16 subjects, 2 weeks apart, were compared to determine reproducibility