Evaluation of hemoglobin A1c measurement by Capillarys 2 electrophoresis for detection of abnormal glucose tolerance in African immigrants to the United States.


ABSTRACT:

Background

Hemoglobin A1c (HbA1c) is used to monitor long-term glycemic control in individuals with diabetes, guide therapy, predict the risk of microvascular complications, and more recently to diagnose diabetes. An automated liquid-flow capillary electrophoresis method was recently developed to measure HbA1c using the Capillarys 2 Flex Piercing instrument.

Methods

Analytical evaluation was performed at 2 clinical centers. A clinical analysis was conducted in 109 African-born individuals, 24% of whom have variant hemoglobin (HbAS or HbAC). Abnormal glucose tolerance (which includes both diabetes and prediabetes) was defined as 2h glucose of ≥ 140 mg/dl (7.8 mmol/l) during an oral glucose tolerance test.

Results

Interlaboratory CVs were ≤ 2.1%. The method showed satis