Rapid point-of-care concentration of bacteria in a disposable microfluidic device using meniscus dragging effect.
Ontology highlight
ABSTRACT: We report a low cost, disposable polymer microfluidic sample preparation device to perform rapid concentration of bacteria from liquid samples using enhanced evaporation targeted at downstream detection using surface enhanced Raman spectroscopy (SERS). The device is composed of a poly(dimethylsiloxane) (PDMS) liquid sample flow layer, a reusable metal airflow layer, and a porous PTFE (Teflon™) membrane sandwiched in between the liquid and air layers. The concentration capacity of the device was successfully demonstrated with fluorescently tagged Escherichia coli (E. coli). The recovery concentration was above 85% for all initial concentrations lower than 1 × 10(4) CFU mL(-1). In the lowest initial concentration cases, 100 µL initial volumes of bacteria solution at 100 CFU mL(-1) were conce