ABSTRACT: Background
Cuff pressure (P cuff) control is mandatory to avoid leakage of oral secretions passing the tracheal tube and tracheal ischemia. The aim of the present trial was to determine the efficacy of a mechanical device (PressureEasy®) in the continuous control of P cuff in patients intubated with polyvinyl chloride (PVC)-cuffed tracheal tubes, compared with routine care using a manometer.Methods
This is a prospective, randomized, controlled, cross-over study. All patients requiring intubation with a predicted duration of mechanical ventilation ≥48 h were eligible. Eighteen patients randomly received continuous control of P cuff with PressureEasy® device for 24 h, followed by discontinuous control (every 4 h) with a manual manometer for 24 h, or vice versa. P cuff and air