Improved Standardization of the Bio-Rad Platelia Aspergillus Galactomannan Antigen Sandwich Enzyme Immunoassay Using the DS2 (Dynex) Enzyme-Linked Immunosorbent Assay (ELISA) Processing System.
Ontology highlight
ABSTRACT: The galactomannan enzyme immunoassay (GM-EIA) is widely utilized for the diagnosis of invasive aspergillosis (IA). There is inconsistent reproducibility of results between centers when the assay is processed manually. Automation of EIAs can reduce variation. This study investigated the semiautomation of the GM-EIA on the DS2 (Dynex) platform in the following three stages: (i) DS2 GM-EIA method validation with experimental samples, (ii) DS2 retesting of case-defined clinical samples, and (iii) a 12-month audit of DS2 GM-EIA performance. In stage i, Bland-Altman analysis demonstrated a reduced variance between optical density index (ODI) values for samples processed on two DS2 platforms (mean difference, -0.02; limits of agreement [LOA], -0.19 to 0.14) compared with the variance between samp
SUBMITTER: Gorton RL
PROVIDER: S-EPMC4473217 | biostudies-literature | 2015 Jul
REPOSITORIES: biostudies-literature