ABSTRACT: Introduction
High short-term failure rates have been reported for a variety of metal-on-metal (MoM) total hip replacements (THRs) owing to adverse reactions to metal debris (ARMD). This has led to the withdrawal of certain poorly performing THRs. This study analysed the outcomes of a MoM THR system.Methods
Between 2004 and 2010, 578 uncemented MoM THRs (511 patients, mean age: 60.0 years) were implanted at one specialist centre. The THR system used consisted of the Corail(®) stem, Pinnacle(®) cup, Ultamet(®) liner and Articul/eze(®) femoral head (all DePuy, Leeds, UK). All patients were recalled for clinical review with imaging performed as necessary.Results
The mean follow-up duration was 5.0 years (range: 1.0-9.1 years). Overall, 39 hips (6.7%) in 38 patients (all