A method based on light scattering to estimate the concentration of virus particles without the need for virus particle standards.


ABSTRACT: Most often the determination of the concentration of virus particles is rendered difficult by the availability of proper standards. We have adapted a static light scattering based method for the quantification of virus particles (shown for poliovirus) without the need of virus particle standards. Instead, as standards, well-characterized polymeric nanoparticle solutions are used. The method is applicable for virus particles acting as Rayleigh scatterers, i.e., virus particles with equivalent diameters up to ca. 1/10th of the wavelength of the scattered monochromatic light (∼70 nm diameter). Further limitations may arise if the refractive index of the virus is unavailable or cannot be calculated based on its composition, such as in case of enveloped viruses. The method is especially relevan

SUBMITTER: Makra I 

PROVIDER: S-EPMC4487338 | biostudies-literature | 2015