A new method of sweat testing: the CF Quantum®sweat test.


ABSTRACT:

Background

Conventional methods of sweat testing are time consuming and have many steps that can and do lead to errors. This study compares conventional sweat testing to a new quantitative method, the CF Quantum® (CFQT) sweat test. This study tests the diagnostic accuracy and analytic validity of the CFQT.

Methods

Previously diagnosed CF patients and patients who required a sweat test for clinical indications were invited to have the CFQT test performed. Both conventional sweat testing and the CFQT were performed bilaterally on the same day. Pairs of data from each test are plotted as a correlation graph and Bland-Altman plot. Sensitivity and specificity were calculated as well as the means and coefficient of variation by test and by extremity. After completing the study, su