Using BD Vacutainer CD4 Stabilization Tubes for Absolute Cluster of Differentiation Type 4 Cell Count Measurement on BD FacsCount and Partec Cyflow Cytometers: A Method Comparison Study from Zimbabwe.


ABSTRACT:

Background

Blood collected in conventional EDTA tubes requires laboratory analysis within 48 hours to provide valid CD4 cell count results. This restricts access to HIV care for patients from rural areas in resource-constraint settings due to sample transportation problems. Stabilization Tubes with extended storage duration have been developed but not yet evaluated comprehensively.

Objective

To investigate stability of absolute CD4 cell count measurement of samples in BD Vacutainer CD4 Stabilization Tubes over the course of 30 days.

Methods

This was a laboratory-based method comparison study conducted at a rural district hospital in Beitbridge, Zimbabwe. Whole peripheral blood from 88 HIV positive adults was drawn into BD Vacutainer CD4 Stabilization Tubes and re-test