First report of 90-day support of 2 calves with a continuous-flow total artificial heart.


ABSTRACT:

Objective

The Cleveland Clinic continuous-flow total artificial heart (CFTAH) is a compact, single-piece, valveless, pulsatile pump providing self-regulated hemodynamic output to left/right circulation. We evaluated chronic in vivo pump performance, physiologic and hemodynamic parameters, and biocompatibility of the CFTAH in a well-established calf model.

Methods

CFTAH pumps have been implanted in 17 calves total. Hemodynamic parameters, pump performance, and device-related adverse events were evaluated during studies and at necropsy.

Results

In vivo experiments demonstrated good hemodynamic performance (pump flow, 7.3 ± 0.7 L/min; left atrial pressure, 16 ± 3 mm Hg; right atrial pressure, 17 ± 3 mm Hg; right atrial pressure-left atrial pressure difference, 1 ± 2 mm H