Unknown

Dataset Information

Ultra-High-Resolution Computed Tomography of the Lung: Image Quality of a Prototype Scanner.


ABSTRACT:

Purpose

The image noise and image quality of a prototype ultra-high-resolution computed tomography (U-HRCT) scanner was evaluated and compared with those of conventional high-resolution CT (C-HRCT) scanners.

Materials and methods

This study was approved by the institutional review board. A U-HRCT scanner prototype with 0.25 mm x 4 rows and operating at 120 mAs was used. The C-HRCT images were obtained using a 0.5 mm x 16 or 0.5 mm x 64 detector-row CT scanner operating at 150 mAs. Images from both scanners were reconstructed at 0.1-mm intervals; the slice thickness was 0.25 mm for the U-HRCT scanner and 0.5 mm for the C-HRCT scanners. For both scanners, the display field of view was 80 mm. The image noise of each scanner was evaluated using a phantom. U-HRCT and C-HRCT image

SUBMITTER: Kakinuma R 

PROVIDER: S-EPMC4564227 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature