ABSTRACT: Aims
To assess the feasibility of high-resolution quantitative cardiovascular magnetic resonance (CMR) voxel-wise perfusion imaging using clinical 1.5 and 3 T sequences and to validate it using fluorescently labelled microspheres in combination with a state of the art imaging cryomicrotome in a novel, isolated blood-perfused MR-compatible free beating pig heart model without respiratory motion.Methods and results
MR perfusion imaging was performed in pig hearts at 1.5 (n = 4) and 3 T (n = 4). Images were acquired at physiological flow ('rest'), reduced flow ('ischaemia'), and during adenosine-induced hyperaemia ('stress') in control and coronary occlusion conditions. Fluorescently labelled microspheres and known coronary myocardial blood flow represented the reference stand