Accuracy Evaluation of CONTOUR(®)PLUS Compared With Four Blood Glucose Monitoring Systems.


ABSTRACT:

Introduction

The aim of this study was to compare the accuracy of 5 blood glucose monitoring systems (BGMSs; CONTOUR(®)PLUS [CP], Accu-Chek(®) Active [ACA], Accu-Chek(®) Performa [ACP], FreeStyle Freedom™ [FF], OneTouch(®) SelectSimple™ [OTSS]).

Methods

Study staff tested fingerstick samples from 106 subjects aged ≥18 years using the 5 BGMSs. Some samples were modified to achieve blood glucose concentrations throughout the measuring range. The primary endpoint was comparison of the mean absolute relative difference (MARD) from the reference value (Yellow Springs Instruments [YSI]) across the overall tested glucose range. Other endpoints were MARD in the low (≤80 mg/dL [≤4.4 mmol/L]), middle (81-180 mg/dL [4.5-10.0 mmol/L]), and high (>180 mg/dL [>10.0 mmol/L]) glucose ranges