25OHD analogues and vacuum blood collection tubes dramatically affect the accuracy of automated immunoassays.


ABSTRACT: Variations in vitamin D quantification methods are large, and influences of vitamin D analogues and blood collection methods have not been systematically examined. We evaluated the effects of vitamin D analogues 25OHD2 and 3-epi 25OHD3 and blood collection methods on vitamin D measurement, using five immunoassay systems and liquid chromatography-tandem mass spectrometry (LC-MS/MS). Serum samples (332) were selected from routine vitamin D assay requests, including samples with or without 25OHD2 or 3-epi 25OHD3, and analysed using various immunoassay systems. In samples with no 25OHD2 or 3-epi 25OHD3, all immunoassays correlated well with LC-MS/MS. However, the Siemens system produced a large positive mean bias of 12.5 ng/mL and a poor Kappa value when using tubes with clot activator and gel