ABSTRACT: Objective
Pulse oximetry, which noninvasively detects the blood flow of peripheral tissue, has achieved widespread clinical use. We have noticed that the better the quality of cardiopulmonary resuscitation (CPR), the better the appearance of pulse oximetry plethysmographic waveform (POP). We investigated whether the area under the curve (AUC) and/or the amplitude (Amp) of POP could be used to monitor the quality of CPR.Design
Prospective, randomized controlled study.Setting
Animal experimental center in Peking Union Medical Collage Hospital, Beijing, China.Subjects
Healthy 3-month-old male domestic swine.Interventions
34 local pigs were enrolled in this study. After 4 minutes of untreated ventricular fibrillation, animals were randomly assigned into