A Digital Reference Object to Analyze Calculation Accuracy of PET Standardized Uptake Value.


ABSTRACT:

Purpose

To determine the extent of variations in computing standardized uptake value (SUV) by body weight (SUV(BW)) among different software packages and to propose a Digital Imaging and Communications in Medicine (DICOM) reference test object to ensure the standardization of SUV computation between medical image viewing workstations.

Materials and methods

Research ethics board approval was not necessary because this study only evaluated images of a phantom. A synthetic set of positron emission tomographic (PET)/computed tomographic (CT) image data, called a digital reference object (DRO), with known SUV was created. The DRO was sent to 16 sites and evaluated on 21 different PET/CT display software packages. Users were asked to draw various regions of interest (ROIs) on spec