A simple and inexpensive method to control oxygen concentrations within physiological and neoplastic ranges.
Ontology highlight
ABSTRACT: Traditional methods for regulating oxygen concentration ([O2]) in in vitro experiments over the range found in normal and tumor tissues require the use of expensive equipment to generate controlled gas atmospheres or the purchase of a range of gas cylinders with certified O2 percentages. Here we describe a simple and inexpensive enzymatic method for generating low, precise steady-state [O2] levels that are stable for several hours. This method is particularly applicable to the in vitro study of some classes of hypoxia-targeted antitumor prodrugs and bioreductively activated agents.
SUBMITTER: Penketh PG
PROVIDER: S-EPMC4631621 | biostudies-literature | 2015 Dec