MALDI-TOF Mass Spectrometry for Multilocus Sequence Typing of Escherichia coli Reveals Diversity among Isolates Carrying blaCMY₋₂-Like Genes.


ABSTRACT: Effective surveillance and management of pathogenic Escherichia coli relies on robust and reproducible typing methods such as multilocus sequence typing (MLST). Typing of E. coli by MLST enables tracking of pathogenic clones that are known to carry virulence factors or spread resistance, such as the globally-prevalent ST131 lineage. Standard MLST for E. coli requires sequencing of seven alleles, or a whole genome, and can take several days. Here, we have developed and validated a nucleic-acid-based MALDI-TOF mass spectrometry (MS) method for MLST as a rapid alternative to sequencing that requires minimal operator expertise. Identification of alleles was 99.6% concordant with sequencing. We employed MLST by MALDI-TOF MS to investigate diversity among 62 E. coli isolates from Sydney, Austral