Ontology highlight
ABSTRACT: Unlabelled
Direct-reading organic vapor monitors are often used to measure volatile organic compound concentrations in complex chemical gas mixtures. However, there is a paucity of data on the impact of multiple gases on monitor performance, even though it is known that monitor sensitivity may vary by chemical. This study investigated the effects of interferents on the performance of the MIRAN SapphIRe Portable Ambient Air Analyzer (SAP) and Century Portable Toxic Vapor Analyzer (TVA-1000) when sampling a specific agent of interest (cyclohexane). The TVA-1000 contained a dual detector: a photoionization detector (PID) and a flame ionization detector (FID). Three devices of each monitor were challenged with different combinations of cyclohexane and potential interferent vapors (hexa
SUBMITTER: LeBouf RF
PROVIDER: S-EPMC4657743 | biostudies-literature | 2015 Mar
REPOSITORIES: biostudies-literature