ABSTRACT: Purpose
To compare and combine dual-energy based and iterative metal artefact reduction on hip prosthesis and dental implants in CT.Material and methods
A total of 46 patients (women:50%,mean age:63±15years) with dental implants or hip prostheses (n = 30/20) were included and examined with a second-generation Dual Source Scanner. 120kV equivalent mixed-images were derived from reconstructions of the 100/Sn140kV source images using no metal artefact reduction (NOMAR) and iterative metal artefact reduction (IMAR). We then generated monoenergetic extrapolations at 130keV from source images without IMAR (DEMAR) or from source images with IMAR, (IMAR+DEMAR). The degree of metal artefact was quantified for NOMAR, IMAR, DEMAR and IMAR+DEMAR using a Fourier-based method and subject