Influence of temporal regularization and radial undersampling factor on compressed sensing reconstruction in dynamic contrast enhanced MRI of the breast.


ABSTRACT:

Background

To evaluate the influence of temporal sparsity regularization and radial undersampling on compressed sensing reconstruction of dynamic contrast-enhanced (DCE) MRI, using the iterative Golden-angle RAdial Sparse Parallel (iGRASP) MRI technique in the setting of breast cancer evaluation.

Methods

DCE-MRI examinations of the breast (n = 7) were conducted using iGRASP at 3 Tesla. Images were reconstructed with five different radial undersampling schemes corresponding to temporal resolutions between 2 and 13.4 s/frame and with four different weights for temporal sparsity regularization (λ = 0.1, 0.5, 2, and 6 times of noise level). Image similarity to time-averaged reference images was assessed by two breast radiologists and using quantitative metrics. Temporal similarity was measured in terms of wash-in slope and contrast kinetic model parameters.

Results

iGRASP images reconstructed with λ = 2 and 5.1 s/frame had significantly (P < 0.05) higher similarity to time-averaged reference images than the images with other reconstruction parameters (mutual information (MI) >5%), in agreement with the assessment of two breast radiologists. Higher undersampling (temporal resolution < 5.1 s/frame) required stronger temporal sparsity regularization (λ ≥ 2) to remove streaking aliasing artifacts (MI > 23% between λ = 2 and 0.5). The difference between the kinetic-model transfer rates of benign and malignant groups decreased as temporal resolution decreased (82% between 2 and 13.4 s/frame).

Conclusion

This study demonstrates objective spatial and temporal similarity measures can be used to assess the influence of sparsity constraint and undersampling in compressed sensing DCE-MRI and also shows that the iGRASP method provides the flexibility of optimizing these reconstruction parameters in the postprocessing stage using the same acquired data.

SUBMITTER: Kim SG 

PROVIDER: S-EPMC4666836 | biostudies-literature | 2016 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Influence of temporal regularization and radial undersampling factor on compressed sensing reconstruction in dynamic contrast enhanced MRI of the breast.

Kim Sungheon G SG   Feng Li L   Grimm Robert R   Freed Melanie M   Block Kai Tobias KT   Sodickson Daniel K DK   Moy Linda L   Otazo Ricardo R  

Journal of magnetic resonance imaging : JMRI 20150601 1


<h4>Background</h4>To evaluate the influence of temporal sparsity regularization and radial undersampling on compressed sensing reconstruction of dynamic contrast-enhanced (DCE) MRI, using the iterative Golden-angle RAdial Sparse Parallel (iGRASP) MRI technique in the setting of breast cancer evaluation.<h4>Methods</h4>DCE-MRI examinations of the breast (n = 7) were conducted using iGRASP at 3 Tesla. Images were reconstructed with five different radial undersampling schemes corresponding to temp  ...[more]

Similar Datasets

| S-EPMC11616894 | biostudies-literature
| S-EPMC10257135 | biostudies-literature
| S-EPMC4583338 | biostudies-literature
| S-EPMC3991777 | biostudies-literature
| S-EPMC6755916 | biostudies-literature
| S-EPMC6126447 | biostudies-literature
| S-EPMC8854616 | biostudies-literature
| S-EPMC9617779 | biostudies-literature
| S-EPMC11460721 | biostudies-literature
| S-EPMC9314035 | biostudies-literature