ABSTRACT: Purpose
The use of new headless compression screws (HCSs) for scaphoid fixation is growing, but the nonunion rate has remained constant. The aim of this study was to compare the stability of fixation resulting from four modern HCSs using a simulated fracture model to determine the optimal screw design(s).Methods
We tested 40 fresh-frozen cadaver scaphoids treated with the Acumed Acutrak 2 mini (AA), the KLS Martin HBS2 midi (MH), the Stryker TwinFix (ST) and the Synthes HCS 3.0 with a long thread (SH). The bones with simulated fractures and implanted screws were loaded uniaxially into flexion for 2000 cycles with a constant bending moment of 800 Nmm. The angulation of the fracture fragments was measured continuously. Data were assessed statistically using the univariate ANO