Spectral fitting using basis set modified by measured B0 field distribution.
Ontology highlight
ABSTRACT: This study sought to demonstrate and evaluate a novel spectral fitting method to improve quantification accuracy in the presence of large magnetic field distortion, especially with high fields. MRS experiments were performed using a point-resolved spectroscopy (PRESS)-type sequence at 7 T. A double-echo gradient echo (GRE) sequence was used to acquire B0 maps following MRS experiments. The basis set was modified based on the measured B0 distribution within the MRS voxel. Quantification results were obtained after fitting the measured MRS data using the modified basis set. The proposed method was validated using numerical Monte Carlo simulations, phantom measurements, and comparison of occipital lobe MRS measurements under homogeneous and inhomogeneous magnetic field conditions. In vivo res
SUBMITTER: Li N
PROVIDER: S-EPMC4715526 | biostudies-literature | 2015 Dec
REPOSITORIES: biostudies-literature
ACCESS DATA