ABSTRACT: Background
Accuracy/robustness of HbA1c estimation (eA1c) with an algorithm built into the MyStar Extra blood glucose (BG) meter has been demonstrated by in silico testing. We evaluated the performance and use of eA1c in a clinical setting.Methods
Subjects took the BG meter home for 4 months to obtain eA1c in this open-label, single-center study. Laboratory HbA1c values were obtained approximately every 2 weeks and the corresponding eA1c documented. Subjects completed a questionnaire at study end (NCT01885546).Results
There were 133 enrolled subjects (mean [SD] age 60.0 [15.0] years, 69 males, 104 with diabetes, HbA1c 7.0% [1.4]). A total of 1008 pairs of eA1c and laboratory HbA1c values were available. In subjects with diabetes, 97.5% of the eA1c results fell withi