Sensitivity of the Predictive Hypoglycemia Minimizer System to the Algorithm Aggressiveness Factor.


ABSTRACT:

Background

The Predictive Hypoglycemia Minimizer System ("Hypo Minimizer"), consisting of a zone model predictive controller (the "controller") and a safety supervision module (the "safety module"), aims to mitigate hypoglycemia by preemptively modulating insulin delivery based on continuous glucose monitor (CGM) measurements. The "aggressiveness factor," a pivotal variable in the system, governs the speed and magnitude of the controller's insulin dosing characteristics in response to changes in CGM levels.

Methods

Twelve adults with type 1 diabetes were studied in closed-loop in a clinical research center for approximately 24 hours. This analysis focused primarily on the effect of the aggressiveness factor on the automated insulin-delivery characteristics of the controller,

SUBMITTER: Finan DA 

PROVIDER: S-EPMC4738202 | biostudies-literature | 2015 Jun