Ontology highlight
ABSTRACT: Background
The Predictive Hypoglycemia Minimizer System ("Hypo Minimizer"), consisting of a zone model predictive controller (the "controller") and a safety supervision module (the "safety module"), aims to mitigate hypoglycemia by preemptively modulating insulin delivery based on continuous glucose monitor (CGM) measurements. The "aggressiveness factor," a pivotal variable in the system, governs the speed and magnitude of the controller's insulin dosing characteristics in response to changes in CGM levels.Methods
Twelve adults with type 1 diabetes were studied in closed-loop in a clinical research center for approximately 24 hours. This analysis focused primarily on the effect of the aggressiveness factor on the automated insulin-delivery characteristics of the controller,
SUBMITTER: Finan DA
PROVIDER: S-EPMC4738202 | biostudies-literature | 2015 Jun
REPOSITORIES: biostudies-literature