ABSTRACT: Objective
Deficient retro-aortic rim has been identified as a risk factor for device erosion following trans-catheter closure of atrial septal defects (ASDs). Transthoracic echocardiography (TTE) is the primary screening method for subjects for possible device closure of ASD, but its reliability in measuring retro-aortic rim size has not been assessed previously.Design
A single-institution cross-sectional analysis of children and adults referred for trans-catheter device closure of single ostium secundum ASD from January 1, 2005 to April 1, 2012 with reviewable TTE and trans-esophageal echocardiogram images was performed. Inter-rater reliability of measurements was tested in a 24% sample. Accuracy of TTE measurement of retro-aortic rim was assessed using a Bland-Altman plot