ABSTRACT: Background
Optical coherence tomography (OCT), can be used in clinical practice to provide high resolution cross-sectional images of the retina, optic disc and macula structure. These measurements can be useful for early detection, diagnosis, monitoring and treatment guidance for retinal diseases. Therefore, repeatability of measurements in OCT is of great importance.Methods
Macula and optic disc parameters from the right eye of 30 healthy subjects were obtained twice with the Canon OCT-HS100 and Zeiss Cirrus HD-OCT 5000. Repeatability was evaluated by use of the coefficient of repeatability (CR) and the coefficient of repeatability as a percentage of the mean (CR%), and the obtained measurements were compared between the instruments.Results
CR% of optic disc parameters ranged between 0.90 and 22.22% and 0.00 and 16.00% with the Canon and Zeiss OCT respectively. For macular parameters CR% ranged between 0.62 and 2.81% and 0.99 and 1.81% with the Canon and Zeiss OCT respectively. No statistical difference could be found when comparing the CR of all macular and disc measurements between the instruments. Compared to our previously published data repeatability has significantly improved with the inclusion of automatic tracking systems with both the Canon and Zeiss OCT.Conclusion
Automatic tracking function improves repeatability in both Canon OCT-HS100 and Zeiss Cirrus HD-OCT 5000. However, measurements generated by the two instruments are still not interchangeable.
SUBMITTER: Brautaset R
PROVIDER: S-EPMC4750906 | biostudies-literature | 2016
REPOSITORIES: biostudies-literature
Brautaset Rune R Birkeldh Ulrika U Frehr Alstig Petra P Wikén Petra P Nilsson Maria M
PloS one 20160211 2
<h4>Background</h4>Optical coherence tomography (OCT), can be used in clinical practice to provide high resolution cross-sectional images of the retina, optic disc and macula structure. These measurements can be useful for early detection, diagnosis, monitoring and treatment guidance for retinal diseases. Therefore, repeatability of measurements in OCT is of great importance.<h4>Methods</h4>Macula and optic disc parameters from the right eye of 30 healthy subjects were obtained twice with the Ca ...[more]