Identification and reduction of image artifacts in non-contrast-enhanced velocity-selective peripheral angiography at 3T.


ABSTRACT:

Purpose

To identify and reduce image artifacts in non-contrast-enhanced velocity-selective (VS) magnetization-prepared peripheral MR angiography (MRA) at 3T.

Methods

To avoid signal loss in the arteries, double and quadruple refocused VS excitation pulse sequences were designed that were robust to a wide range of B0 and B1 offset. To suppress stripe artifact and background signal variation, we successively applied two VS preparations with excitation profiles shifted by half the period of the stripes. VS-MRA using single, double, and quadruple refocused VS preparations was tested in healthy subjects and a patient.

Results

In the regions of large B0 and B1 offsets, arterial signal loss was yielded by single refocused VS preparation, but was avoided with double or quadru