Comparison of Dixon Sequences for Estimation of Percent Breast Fibroglandular Tissue.


ABSTRACT:

Objectives

To evaluate sources of error in the Magnetic Resonance Imaging (MRI) measurement of percent fibroglandular tissue (%FGT) using two-point Dixon sequences for fat-water separation.

Methods

Ten female volunteers (median age: 31 yrs, range: 23-50 yrs) gave informed consent following Research Ethics Committee approval. Each volunteer was scanned twice following repositioning to enable an estimation of measurement repeatability from high-resolution gradient-echo (GRE) proton-density (PD)-weighted Dixon sequences. Differences in measures of %FGT attributable to resolution, T1 weighting and sequence type were assessed by comparison of this Dixon sequence with low-resolution GRE PD-weighted Dixon data, and against gradient-echo (GRE) or spin-echo (SE) based T1-weighted Dixon datasets, respectively.

Results

%FGT measurement from high-resolution PD-weighted Dixon sequences had a coefficient of repeatability of ±4.3%. There was no significant difference in %FGT between high-resolution and low-resolution PD-weighted data. Values of %FGT from GRE and SE T1-weighted data were strongly correlated with that derived from PD-weighted data (r = 0.995 and 0.96, respectively). However, both sequences exhibited higher mean %FGT by 2.9% (p < 0.0001) and 12.6% (p < 0.0001), respectively, in comparison with PD-weighted data; the increase in %FGT from the SE T1-weighted sequence was significantly larger at lower breast densities.

Conclusion

Although measurement of %FGT at low resolution is feasible, T1 weighting and sequence type impact on the accuracy of Dixon-based %FGT measurements; Dixon MRI protocols for %FGT measurement should be carefully considered, particularly for longitudinal or multi-centre studies.

SUBMITTER: Ledger AE 

PROVIDER: S-EPMC4806997 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of Dixon Sequences for Estimation of Percent Breast Fibroglandular Tissue.

Ledger Araminta E W AE   Scurr Erica D ED   Hughes Julie J   Macdonald Alison A   Wallace Toni T   Thomas Karen K   Wilson Robin R   Leach Martin O MO   Schmidt Maria A MA  

PloS one 20160324 3


<h4>Objectives</h4>To evaluate sources of error in the Magnetic Resonance Imaging (MRI) measurement of percent fibroglandular tissue (%FGT) using two-point Dixon sequences for fat-water separation.<h4>Methods</h4>Ten female volunteers (median age: 31 yrs, range: 23-50 yrs) gave informed consent following Research Ethics Committee approval. Each volunteer was scanned twice following repositioning to enable an estimation of measurement repeatability from high-resolution gradient-echo (GRE) proton-  ...[more]

Similar Datasets

| S-EPMC8515131 | biostudies-literature
| S-EPMC10468506 | biostudies-literature
| S-EPMC7547401 | biostudies-literature
| S-EPMC3852242 | biostudies-other
| S-EPMC11256640 | biostudies-literature
| S-EPMC10628070 | biostudies-literature
| S-EPMC4123610 | biostudies-literature
| S-EPMC10044359 | biostudies-literature
| S-EPMC4596769 | biostudies-literature
| S-EPMC5429378 | biostudies-literature