ABSTRACT: Purpose
To compare visualization of choroidal neovascularization (CNV) secondary to age-related macular degeneration (AMD) using an ultrahigh-speed swept-source (SS) optical coherence tomography angiography (OCTA) prototype vs a spectral-domain (SD) OCTA device.Design
Comparative analysis of diagnostic instruments.Methods
Patients were prospectively recruited to be imaged on SD OCT and SS OCT devices on the same day. The SD OCT device employed is the RTVue Avanti (Optovue, Inc, Fremont, California, USA), which operates at ∼840 nm wavelength and 70 000 A-scans/second. The SS OCT device used is an ultrahigh-speed long-wavelength prototype that operates at ∼1050 nm wavelength and 400 000 A-scans/second. Two observers independently measured the CNV area on OCTA en face