Use of captive spray ionization to increase throughput of the data-independent acquisition technique PAcIFIC.


ABSTRACT:

Rationale

The Precursor Acquisition Independent From Ion Count (PAcIFIC) method is a data-independent acquisition technique capable of identifying proteins over eight orders of magnitude in a single analysis in human plasma. Widespread application of this technique in proteomic studies is hindered by its time-intensive nature. There exists a need to explore strategies to increase the throughput of the PAcIFIC method.

Methods

The PAcIFIC acquisition technique was optimized for use with an Orbitrap mass spectrometer fitted with a captive spray ionization (CSI) source. Chromatographic methods, PAcIFIC acquisition parameters, for example, channels interrogated per chromatographic gradient, time span of chromatographic gradient, and sample loading amount, were investigated to achieve a maximum number of peptide and protein identifications on a yeast proteome where protein copy number had been previously determined.

Results

A 24-hour CSI PAcIFIC method was developed with minimal reduction of peptide and protein identifications from the 4.2-day nano-electrospray ionization (nESI) PAcIFIC method. Analysis of a yeast cell lysate with the 4.2-day nESI PAcIFIC method resulted in 13,468 peptide and 2120 protein identifications. A 24-hour CSI PAcIFIC method resulted in 11,277 peptide and 1753 protein identifications. Increased sample loading of the CSI system allowed for an 8% increase in peptide and protein identifications.

Conclusions

A dramatic decrease in the overall analysis time of the PAcIFIC method (24 h with CSI versus 100.8 h with nESI) was achieved with minimal reduction of peptide and protein identifications. Furthermore, the CSI PAcIFIC method demonstrated a high degree of sensitivity and capability of identifying proteins across a large dynamic range observed with the nESI PAcIFIC method (CSI PAcIFIC identified proteins as low as 46 molecules per cell).

SUBMITTER: Chapman JD 

PROVIDER: S-EPMC4830633 | biostudies-literature | 2016 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Use of captive spray ionization to increase throughput of the data-independent acquisition technique PAcIFIC.

Chapman John D JD   Edgar J Scott JS   Goodlett David R DR   Goo Young Ah YA  

Rapid communications in mass spectrometry : RCM 20160501 9


<h4>Rationale</h4>The Precursor Acquisition Independent From Ion Count (PAcIFIC) method is a data-independent acquisition technique capable of identifying proteins over eight orders of magnitude in a single analysis in human plasma. Widespread application of this technique in proteomic studies is hindered by its time-intensive nature. There exists a need to explore strategies to increase the throughput of the PAcIFIC method.<h4>Methods</h4>The PAcIFIC acquisition technique was optimized for use  ...[more]

Similar Datasets

| S-EPMC9024139 | biostudies-literature
| S-EPMC12765230 | biostudies-literature
| S-EPMC8151837 | biostudies-literature
| S-EPMC3591489 | biostudies-literature
| S-EPMC6537876 | biostudies-literature