Comparison of GC-MS and GC×GC-MS in the analysis of human serum samples for biomarker discovery.
Ontology highlight
ABSTRACT: We compared the performance of gas chromatography time-of-flight mass spectrometry (GC-MS) and comprehensive two-dimensional gas chromatography mass spectrometry (GC×GC-MS) for metabolite biomarker discovery. Metabolite extracts from 109 human serum samples were analyzed on both platforms with a pooled serum sample analyzed after every 9 biological samples for the purpose of quality control (QC). The experimental data derived from the pooled QC samples showed that the GC×GC-MS platform detected about three times as many peaks as the GC-MS platform at a signal-to-noise ratio SNR ≥ 50, and three times the number of metabolites were identified by mass spectrum matching with a spectral similarity score Rsim ≥ 600. Twenty-three metabolites had statistically significant abundance changes between