Positive Contrast MRI Techniques for Visualization of Iron-Loaded Hernia Mesh Implants in Patients.


ABSTRACT:

Object

In MRI, implants and devices can be delineated via susceptibility artefacts. To discriminate susceptibility voids from proton-free structures, different positive contrast techniques were implemented. The purpose of this study was to evaluate a pulse sequence-based positive contrast technique (PCSI) and a post-processing susceptibility gradient mapping algorithm (SGM) for visualization of iron loaded mesh implants in patients.

Material and methods

Five patients with iron-loaded MR-visible inguinal hernia mesh implants were examined at 1.5 Tesla. A gradient echo sequence (GRE; parameters: TR: 8.3ms; TE: 4.3ms; NSA:2; FA:20°; FOV:350mm²) and a PCSI sequence (parameters: TR: 25ms; TE: 4.6ms; NSA:4; FA:20°; FOV:350mm²) with on-resonant proton suppression were performed. SG