Comparison of LC-MS-MS and GC-MS Analysis of Benzodiazepine Compounds Included in the Drug Demand Reduction Urinalysis Program.


ABSTRACT: Liquid chromatography-tandem mass spectrometry (LC-MS-MS) offers specific advantages over gas chromatography-mass spectrometry (GC-MS) such as the ability to identify and measure a broader range of compounds with minimal sample preparation. Comparative analysis of LC-MS-MS versus GC-MS was performed for urinalysis detection of five benzodiazepine compounds currently part of the Department of Defense (DoD) Drug Demand Reduction Program (DDRP) testing panel; alpha-hydroxyalprazolam, oxazepam, lorazepam, nordiazepam and temazepam. In the analyses of internally prepared control urine samples at concentrations around the DDRP administrative decision point for benzodiazepines (100 ng/mL), both technologies produced comparable results with average accuracies between 99.7 and 107.3% and average co