Damage in total knee replacements from mechanical overload.


ABSTRACT: The mechanical loads acting across the knee joint following total knee replacements (TKR) during activities of daily living have recently been measured using instrumented TKRs. Using a series of postmortem retrieved TKR constructs we investigated whether these mechanical loads could result in damage to the implant bone interface or supporting bone in the tibia. Eighteen cemented en bloc tibial components (0 to 22 years in service) were loaded under axial compression in increments from 1 to 10 times body weight and digital image correlation was used to measure bone strain and interface micromotion during loading and unloading. Failure was considered to occur when micromotion exceeded 150µm or compressive bone strain exceeded 7300με. The results show that all retrieved specimens had sufficie