Ontology highlight
ABSTRACT: Purpose
To compare grading results between short-wavelength reduced-illuminance and conventional autofluorescence imaging in Stargardt macular dystrophy.Design
Reliability study.Methods
setting: Moorfields Eye Hospital, London (United Kingdom).Patients
Eighteen patients (18 eyes) with Stargardt macular dystrophy.Observation procedures
A series of 3 fundus autofluorescence images using 3 different acquisition parameters on a custom-patched device were obtained: (1) 25% laser power and total sensitivity 87; (2) 25% laser power and freely adjusted sensitivity; and (3) 100% laser power and freely adjusted total sensitivity (conventional). The total area of 2 hypoautofluorescent lesion types (definitely decreased autofluorescence and poorly demarcated que
SUBMITTER: Strauss RW
PROVIDER: S-EPMC4977015 | biostudies-literature | 2016 Aug
REPOSITORIES: biostudies-literature