Automated oxygen titration and weaning with FreeO2 in patients with acute exacerbation of COPD: a pilot randomized trial.


ABSTRACT:

Introduction

We developed a device (FreeO2) that automatically adjusts the oxygen flow rates based on patients' needs, in order to limit hyperoxia and hypoxemia and to automatically wean them from oxygen.

Objective

The aim of this study was to evaluate the feasibility of using FreeO2 in patients hospitalized in the respiratory ward for an acute exacerbation of COPD.

Methods

We conducted a randomized controlled trial comparing FreeO2 vs manual oxygen titration in the respiratory ward of a university hospital. We measured the perception of appropriateness of oxygen titration and monitoring in both groups by nurses and attending physicians using a Likert scale. We evaluated the time in the target range of oxygen saturation (SpO2) as defined for each patient by the attend