Unknown

Dataset Information

Is Model-based Radiostereometric Analysis Suitable for Clinical Trials of a Cementless Tapered Wedge Femoral Stem?


ABSTRACT:

Background

In clinical trials of THA, model-based radiostereometric analysis (RSA) techniques may be less precise than conventional marker-based RSA for measurement of femoral stem rotation. We verified the accuracy and clinical precision of RSA based on computer-aided design models of a cementless tapered wedge femoral stem.

Questions

We asked: (1) Is the accuracy of model-based RSA comparable to that of marker-based RSA? (2) What is the clinical precision of model-based RSA?

Methods

Model-based RSA was performed using combined three-dimensional computer-aided design models of the stem and head provided by the implant manufacturer. The accuracy of model-based RSA was compared with that of marker-based RSA in a phantom model using micromanipulators for controlled tran

SUBMITTER: Nazari-Farsani S 

PROVIDER: S-EPMC5014820 | biostudies-literature | 2016 Oct

REPOSITORIES: biostudies-literature