Evaluation of the Aptima HIV-1 Quant Dx Assay Using Plasma and Dried Blood Spots.
Ontology highlight
ABSTRACT: HIV-1 RNA quantitation in plasma, or virus load testing, is the primary method by which the response to antiretroviral therapy is monitored. Here we describe evaluation of the Aptima HIV-1 Quant Dx assay (Aptima) performed on the automated Panther system. The clinical performance of Aptima was compared to that of the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test v2.0 (CAP/CTM) using 162 EDTA plasma samples collected from patients undergoing HIV-1 monitoring. Overall agreement was 84.0% (136/162), with a kappa statistic of 0.723 (standard error, 0.047; 95% confidence interval [CI], 0.630 to 0.815), indicating substantial agreement. Using the 86 clinical samples quantifiable by both methods, Passing-Bablok regression revealed a regression line of Y = (1.069 × X) - 0.346 (95% CI of the slope [1.003